Table 5: (Part of table) Menu options - Settings (Output)
Options
Description
(If applicable)
To select and set up values for the "Span Check"
output. Example output cycle:
Additional data (Table 6): Select Settings ( )
To select and set up values for the "% Step" output.
% Step
Example output cycle:
Additional data (Table 6): Select Settings ( )
To select and set up values for the "Defined Step"
... Step
output. Example output cycle:
Additional data (Table 6): Select Settings ( )
To select and set up values for the "Ramp" output.
Example output cycle:
Additional data (Table 6): Select Settings ( )
Table 6: Additional data for Settings (Output):
Item
Value
Span Check
Low (0%)
Set the 0% value.
High (100%)
Set the 100% value.
Dwell (d)
Set the period (Hours:Minutes:Seconds) between each
change in value.
% Step
Low (0%), High (100%), Dwell (d): As above.
Step Size (s) ...
Set the change in value for each step as a percentage
%
of the full-scale range (High - Low).
Defined Step
Low (0%), High (100%), Dwell (d): As above.
Step Size (s)
Set the change in value for each step.
Example: 1.000 mA steps.
Ramp
Low (0%), High (100%), Dwell (d): As above.
Travel (t)
Set the period (Hours:Minutes:Seconds) to go from the
Low (0%) value to the High (100%) value.
Auto Repeat
If applicable, select this item to repeat a cycle
continuously.
K0405 Revision B

Operation

Electrical connections

This cycle repeats
automatically.

Communications port connections

Auto Repeat - Optional

Change the output values

Auto Repeat - Optional
Auto Repeat - Optional

Measure/supply mA

This section gives examples of how to connect and use the
instrument. Before you start:
• Read and understand the "Safety" section.
• Do not use a damaged instrument.
To prevent instrument errors, make sure that the electrical
connections (A1-item 10 and/or A2) are correct.
The Help button (A3 - Item 25) shows a connection
diagram for the task selections you have set.
Use the communications port (A1 - item 9) to attach an IDOS
Universal Measurement Module (UMM).
When you attach the cable from a UMM (Figure 13/14), the
instrument automatically changes the menus to give you all
the applicable options (Table 2/3).
When the output operation is set up (Table 5), use one of
these procedures to change the output values:
Table 7: Procedures to change the output
Output
Procedure
Select Edit ( ) and/or use the   buttons. See
the example below.
Select Start/Stop ( ) or use the   buttons to
,
make the step changes manually.
Select Start/Stop ( ).
Example procedure ("Nudge" output):
Display:
1
Edit
mA output
 
Sk1 = Edit
Sk2 = Settings
Display:
5
mA output
2.000
Increment = 1.000
To measure/supply a current:
1. Connect the instrument (Figure 1, 2 or 3) and, if
necessary, adjust the Set Up (Table 1).
2. Select the task from the task selection menu (Table 2/3).
Note: Use the Dual Function area (
the same time. If the area
 selection is not necessary, set
this area to off (). This saves the battery power.
3. If necessary, adjust the Settings (Table 4/5) and/or the
2
3
Edit
+01.000
 
6
3.000
(Table 5)
 ) to do two operations at
[EN] English -