Accessing a parameter
1.
Press PROGRAM
PROGRAM mode.
2.
Either use the ARROW keys
different parameter, or:
3.
Press DISPLAY
field. (The current parameter value remains visible.)
4.
Key in the desired parameter number followed by
ENTER
.
For example: press
5.
The LCD displays the new parameter number and
value.
Changing a Parameter Value
Notes:
• Security must be disabled to enable programming: set P000 to the Unlocked Value stored
in P069. (A remote master can still change configuration, if P799 is set to allow this.)
• Invalid entries will be rejected or limited.
• CLEAR
can be used to clear the field.
1.
Key in the new value.
2.
Press ENTER
to set the value.
Parameter Reset to Factory Default
1.
Scroll to the parameter or enter its address.
2.
Press CLEAR
then ENTER
Master Reset (P999)
Returns all parameters except P000, P069, and P838 to default settings. (The learned TVT curve
is not lost.)
1.
Press PROGRAM
2.
Press DISPLAY
3.
Key in 999.
4.
Press CLEAR
then ENTER
reset. The LCD displays C.ALL.
5.
Reset complete. (Reset takes several seconds to complete.)