D D i i s s i i n n f f e e c c t t i i n n g g
W W A A R R N N I I N N G G
• Always check the CPR connector before use. Replace if visibly soiled. Direct skin contact with soiled material may
increase the risk of infection.
• Always inflate the product and unplug the pump from the mains before you clean or disinfect.
• Do not immerse the product in cleaning or disinfectant solutions.
• Always disinfect the support surface between patients, to avoid the risk of cross-contamination and infection.
• Always make sure that you wipe each product with clean water and dry each product after cleaning or disinfecting. Some
cleaning and disinfecting agents are corrosive in nature and may cause damage to the product if you use them
improperly. If you do not rinse and dry the product, a corrosive residue may be left on the surface of the product that
could cause premature degradation of critical components. Failure to follow these cleaning or disinfecting instructions
may void your warranty.
• Do not allow fluid to pool on the support surface. Fluids may cause premature degradation of components and may
cause unpredictable safety and performance of this product.
C C A A U U T T I I O O N N
• Always completely dry the support surface covers before storing, adding linens, or placing a patient on the surface.
• Do not overexpose the product to higher concentration chemical solutions to avoid the risk of product degradation.
• Always follow the manufacturing instructions, failure may affect the useful life of the product.
Suggested disinfectants:
•
Quaternaries (active ingredient - ammonium chloride) that contain less than 3% glycol ether
•
Phenolic solution (Matar)
•
Chlorinated bleach solution (6500 ppm)
•
Potassium peroxymonosulfate
•
Accelerated hydrogen peroxide
•
70% isopropyl alcohol
To disinfect the support surface covers after each patient use, follow these steps in order:
1. Inflate the support surface with the pump.
2. Disconnect the mains power from the receptacle.
N N o o t t e e - - Leave the hose connected to the support surface to retain the air.
3. Clean and dry the support surface covers before disinfectants are applied.
4. Apply recommended disinfectant solution with a damp cloth or pre-soaked wipes. Do not soak the support surface.
N N o o t t e e - - Make sure that you follow the disinfectant manufacturer's instructions for appropriate contact time and rinsing
requirements.
5. Rinse the product with a clean, damp cloth to remove excess disinfectant.
6. Dry the product with a clean, dry cloth to remove any excess liquid or disinfectant.
7. Allow the product to dry completely before you return to service.