Instructions for Use Vyntus WALK
Ending the measurement - "Standalone" Mode
Page 34/332
All manuals and user guides at all-guides.com
In "Standalone" Mode, a PDF-report will automatically be created after the
measurement:
Tapping on <Print> will automatically save the PDF-report in the"Report
Network Folder" defined in the settings menu (see chapter ""Standalone"
Mode Settings") and will display it in a compressed form. As soon as a Bluetooth
connection has been established with an active printing device, the PDF-report can
be sent for printing.
Tapping on <Finish> will open following window:
Tap on <Yes> to print and/or save the report.
Tapping on <No> will close the 6-Minute Walk Test and the app will automatically
return to the startup-screen WITHOUT the report being saved.
The report header can be
customised (see chapter „Settings
in "Standalone" Mode").
Patient data, BODE-Index as well as
both predicted and actual values of
the 6-Minute Walk Test.
Study conditions, test phase
analysis and values measured in
Baseline, Test and Recovery phase
displayed next to each other.
Graphical overview of the test
phase.
Time of laps, pauses, events and
comments entered.
Version 05.00 • 06-06-2017