Guidance and Manufacturer's Declaration – Electromagnetic Immunity
This device is intended for use in the electromagnetic environment specified below� The user of this device should assure that it
is used in such an environment�
Immunity Test
IEC 60601 Test Level
Conducted RF
3 V
IEC 61000-4-6
150 kHz to 80 MHz
6 V in ISM & amateur
radio bands between
150kHz and 80 MHz
Radiated RF
10 V/m
IEC 61000-4-3
80 MHz to 2�7 GHz
Up to 28 V/m in
telecommunications
bands as specified
in clause 8�10 of
IEC 60601-1-2:2014
Compliance
Level
Electromagnetic Environment - Guidance
Portable and mobile RF communications equipment should be
used no closer to any part of the Device, including cables, than the
recommended separation distance�
Recommended separation distance
3 V
Minimum 30 cm (12 inches)
6 V
Interference may occur in the vicinity of equipment marked
with the following symbol:
10 V/m
28 V/m
SimplyGo User Manual
47