Warning: It is very important to set your device to your prescribed level of
oxygen flow� Do not increase or decrease your flow rate from your
prescribed level until you first consult with your physician or your
healthcare professional�
5�
Position the nasal cannula on your face, as shown below, and breathe
normally through your nose�
To verify the alarm system:
1�
Ensure the device is in Continuous mode� Induce a No Flow Alarm
condition by turning the device ON and completely blocking the patient
cannula connector of the device�
2�
Wait approximately 45 seconds or until the No Flow Alarm occurs�
3�
Verify that the No Flow Alarm indicators and screen symbols are present�
4�
Unblock the patient cannula connector for a minimum of 45 seconds�
5�
Verify that the No Flow Alarm ceases�
Turning the SimplyGo Device Off
1�
When you are ready to turn your SimplyGo device off, press the Power
button� The screen below appears�
2�
Press the Power button a second time and the SimplyGo device will turn
off�
3�
Rather than pressing the Power button a second time, the backlight
intensity may be increased or decreased by repeatedly pressing the Plus
(+) or Minus (-) key until the desired backlight intensity is achieved� The
device may then be powered off using steps 1 and 2 above�
SimplyGo User Manual
21