Tabla de contenido
Idiomas disponibles

Idiomas disponibles

2. Safety / 3. Specifications
15. Protection against external impact
Avoid any kind of external impact. External impacts can generate sparks through
friction processes between different materials.
16. Instrument filling
The filling/refilling of instruments leads to a loss of the explosion protection and
can lead to damage to the instrument.
FR

3. Specifications

Materials (wetted)
Capsule element
Measuring chamber
Process connection
Sealing
Pressure limitation
Steady:
Fluctuating:
Short time:
Temperature effect
When the temperature of the measuring system deviates from the reference
temperature (+20 °C):
max. ±0.6 %/10 K of full scale value
Case ingress protection
IP54
For further specifications, see WIKA data sheet PM 06.06 and the order
documentation.
1) For general use, no ATEX requirement
14
Stainless steel 1.4571 (316 Ti)
Stainless steel 1.4571 (316 Ti)
Stainless steel 1.4571 (316 Ti)
PTFE
Full scale value
0.9 x full scale value
50 x full scale value
(per IEC/EN 60529)
1)
WIKA mode d'emploi manomètre, type 6 selon ATEX
Tabla de contenido
loading

Este manual también es adecuado para:

632.51

Tabla de contenido