How To Respond To Freestyle's Alarm/Light Indicators - AirSep FreeStyle Manual De Instrucciones

Ocultar thumbs Ver también para FreeStyle:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 92

How to Respond to FreeStyle's Alarm/Light Indicators

Status
Indicator
Indicator
Indicator
Indicator
Alarm
Alarm
Alarm
Alarm
MN131-1 rev C
Audible Alarm
Light
Brief, continuous
(Green) pulse;
at start–up
continuous light
No
(Green) pulse;
intermittent light
No
Level indicator
No
(Yellow)
continuous light
Continuous: Beep
(Red) alarm;
continuous light
Intermittent:
25% (Yellow)
Beep, beep, beep.
BATT; battery
indicator.
Rapid intermittent:
(Red) alarm;
Beep, beep, beep
intermittent light
Rapid intermittent:
(Red) alarm;
Beep, beep, beep
continuous light
Indicates
Action
FreeStyle has been
You may begin to operate
turned on
your FreeStyle unit.
FreeStyle is
Continue using FreeStyle
delivering oxygen as
normally.
a pulse flow.
Battery charge level.
Charge as indicated.
Service required.
Return unit to Equipment
Provider for
inspection/service.
Check the cannula
No breath detected
connection. Ensure that you
by the unit for a
are breathing through your
predetermined time
nose. If the alarm persists,
period.
contact your Equipment
provider.
Battery voltage is
Connect the FreeStyle unit
too low to operate
into a DC outlet or an AC
FreeStyle.
outlet immediately.
Breathing rate is
Reduce activity, then if
exceeding the
necessary use another source
capacity of the
of oxygen as available.
FreeStyle unit.
Contact your Equipment
Provider.
General malfunction
Turn off the unit. Change to
of the FreeStyle unit
another source of oxygen,
has occurred.
and contact your
EN-28
Tabla de contenido
loading

Tabla de contenido