3. Screw the irrigation adapter onto the argon gas connection
of the cable.
4. Fill up the syringe completely with irrigation liquid.
5. Screw the syringe onto the luer-lock adapter.
6. Flush the internal lumen of the product to remove as much
contamination as possible.
7. Remove the syringe. Leave the luer-lock adapter on the
product for the next reprocessing steps.
5.6

Manual Cleaning and Disinfection

Use a liquid detergent suitable for the preparation of a dipping
bath.
Use a disinfectant compatible with the detergent suitable for the
preparation of a dipping bath.
The detergent and disinfectant must be suitable for medical de-
vices made of plastic and metal and have a pH value between 5.5
and 11.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of Validated
Procedures".
1. Prepare dipping bath, soak product:
– Use the cleaning solution to flush the internal lumen
several times.
– Fill the internal lumen with cleaning solution.
– Remove the syringe and luer-lock adapter.
– Submerge the product into the cleaning solution com-
pletely. The immersion time must be at least 5 minutes.
– Take the product out of the dipping bath and extend the
cable as far as possible for the next work step.
39