7.3
Required aids
Processing step
Items Needed
Precleaning
Soft plastic brush/soft cloth
Manual cleaning/
Soft plastic brush/soft disposable cloth
disinfection
(low particle)
Compressed air gun (2 bar maximum)
7.4
Precleaning
Use water for precleaning, a non-fixing disinfectant if necessary.
1. Make sure that the probe coupling is securely closed using
the protective cap.
2. Remove surface contaminants with a soft brush/cloth while
holding the product in a water bath and/or rinsing the prod-
uct under running water.
7.5
Manual cleaning and disinfection
Use a liquid detergent suitable for the preparation of a dipping
bath.
Use a disinfectant compatible with the detergent suitable for the
preparation of a dipping bath.
The detergent and disinfectant must be suitable for medical de-
vices made of plastic and metal and have a pH value between 5.5
and 11.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of validated
procedures".