Manual Cleaning And Disinfection - Erbe 20188-300 Manual Del Usuario

Interruptor de pedal para el sistema vio
Tabla de contenido
OBJ_DOKU-186956-008.fm Seite 13 Donnerstag, 26. November 2020 2:59 14
Footswitch 20189-300, 20189-302:
• Yellow pedal: activates cutting
• Blue pedal: activates coagulation
Footswitch 20189-301, 20189-303:
• Yellow pedal: activates cutting
• Blue pedal: activates coagulation
• Black ReMode button: activates the switch between two set-
ting options in a program
Footswitch 20322-101:
• Black pedal: activates smoke evacuation.
5
Cleaning and disinfection
5.1
Safety instructions for processing
Do not exceed a temperature of max. 75 °C!
Erbe recommends wiping or spray disinfection with an appropri-
ate disinfectant. Comply with the disinfectant manufacturer's
stipulations as to its suitability and concentration.
Do not sterilize!
5.2

Manual cleaning and disinfection

Rinse the product with water immediately after use and wash it
with an alkaline cleaning solution.
Disinfect the product after cleaning. When doing so, always
check the instructions provided by the disinfectant manufacturer
to ensure that the agent is suitable and of the correct concentra-
tion.
Dry the product at a temperature of max. 75 °C.