4. If a probe is not already connected to the data logger, connect it now:
• To connect a probe directly to the data logger, first align the orientation mark on the
probe with the line above the probe connector. Then push the probe in all the way, do
not rotate.
• To connect a probe using an extension cable, use an accessory cable that has been
designed for RFL100. Insert the cable connector in the same way as a probe, and then
connect the probe to the cable. Note that HMP110 and HMP110T probes must always
be connected using a cable as they do not mechanically lock into the data logger
when the battery cover is closed.
5. Move the power switch to the On position.
6. Close the battery cover of the data logger. Push the latch down until you hear a click. If
the cover does not close easily, push the probe (or extension cable) in and try again.
7. Look at the display and wait for the data logger to start up. Then verify the status:
• Data logger shows the word NEW on each affected channel when it detects that a
probe has been changed
display shows the measurement reading(s).
Figure 1 Detection of a new RH + T probe
• If dashes "- - -" are shown instead of measurement readings, check that the probe is
properly connected. It is possible to disconnect the probe or cable by pulling it with
the battery cover open.
• Battery indicator should show full batteries
• There should be no error codes shown. If there are, see
1)
This feature is included in RFL100 firmware version 1.2.0 and newer.
6
1)
. The text is shown for a few seconds, after which the
.
Error codes (page
23).
M211822EN-G