Setting up RFL100
Start the installation of RFL100 by performing the appropriate setup procedure:
• If you are connecting 1 humidity or temperature probe, see
temperature probe (page
• If you are connecting 2 temperature probes using the probe splitter accessory, see
with 2 temperature probes (page
• If you are connecting a carbon dioxide (CO
(page
9).
CAUTION!
connecting RFL100 data loggers. Setup instructions include additional firmware
and hardware requirements for using RFL100 with 2 temperature probes, or a
CO
probe. Make sure the components of your viewLinc Monitoring System are
2
updated to the required levels.
When you switch on an RFL100 that has not been accepted to any viewLinc
system yet, it starts to scan for VaiNet access points that are in installation mode.
RFL100 will connect to the access point with the best signal strength, and wait to
be accepted by the administrator of the viewLinc Enterprise Server. Make sure the
access point(s) are installed and in installation mode before starting to set up
RFL100 data loggers.
Setup with 1 humidity or temperature probe
1
2
1. If any cable is connected to the service port of the data logger, disconnect it.
2. Open the battery cover of the data logger.
3. Make sure the power switch is in the Off position.
5).
7).
2
At least viewLinc Enterprise Server version 5.0 is required for
3
Setup with 1 humidity or
) probe, see
Setup with CO₂ probe
1
Power switch.
2
Transport protection cap. Remove
after installation is complete.
3
Latch of the battery cover.
Setup
5