• GMP251 CO
• Probe splitter M8/M12 (item code CBL211050)
• DC power supply with micro-USB connector
• Optional: any supported humidity or temperature probe
• Optional: probe extension cable(s)
CAUTION!
must have firmware version 1.4.0 or higher, and must be manufactured after July
2021. Additionally, make sure all AP10 access points at the location of use have
firmware version 4.0.0.
External power supply is a requirement for operating the data logger with a CO
probe. Batteries are used as a backup power source if the external power fails.
Lithium batteries are recommended as they provide a longer backup time, but
even with lithium batteries the operation time is limited in CO
least 12 hours). The data logger will show the error code Err 106 if it operates in
CO
mode without a power supply.
2
1. If any cable is connected to the service port of the data logger, disconnect it.
2. Open the battery cover of the data logger.
3. Make sure the power switch is in the Off position.
4. Verify that the data logger has lithium batteries (type FR6) inserted. Replace the batteries
if necessary.
5. If the probe splitter M8/M12 is not yet connected to the data logger, connect it now:
a. If a probe or cable is currently connected to the probe connector, pull it straight out
without rotating it.
b. Align the orientation mark on the probe splitter with the line above the probe
connector. Push the probe splitter straight in all the way, do not rotate.
You can also use a probe extension cable between the data logger and
the probe splitter.
6. Connect the CO
2
extension cable if necessary.
7. Optional: connect the temperature or humidity probe to the M8 connector of the probe
splitter, using a suitable extension cable if necessary. If you do not connect a probe to the
M8 connector, make sure the connector is not inserted inside an incubator, for example.
8. Connect the micro-USB connector of the power supply to the service port of the data
logger, and connect the power supply to a wall socket.
9. Move the power switch to the On position.
10
probe
2
To support operation with a CO
probe to the M12 connector of the probe splitter, using a suitable
probe, the RFL100 data logger
2
mode (typically at
2
2
M211822EN-G