1. Product Description
With its broadband spectrum, VALO Cordless is designed to polymerize all light cured products in the wavelength range of 385-515nm per ISO 10650. The VALO Cordless curing light uses
Ultradent VALO rechargeable batteries and battery charger. The curing light is designed to rest in a standard dental unit bracket or can be custom mounted using the bracket included
with the kit.
VALO Cordless Product Components:
• 1 – VALO Cordless curing light
• 4 – Ultradent VALO rechargeable batteries
• 1 – Ultradent VALO battery charger with medical grade 12VDC AC power adapter
• 1 – VALO Barrier Sleeve sample pack
• 1 – VALO Cordless Light Shield
• 1 – Curing light surface mounting bracket with double stick adhesive tape
Overview of Controls:
For all products described, carefully read and understand all instructions and SDS information prior to use.
2. Indications for Use/Intended Purpose
The source of illumination for curing photo-activated dental restorative materials and adhesives.
3. Warnings and Precautions
CAUTION UV emitted from this product. Eye or skin irritation may result from exposure. Use appropriate shielding.
CAUTION Possibly hazardous optical radiation emitted from this product. Do not stare at operating lamp. May be harmful to the eyes.
• DO NOT look directly into the light output. Patient, clinician, and assistants should always wear amber colored UV eye protection when curing light is in use.
• To prevent the risk of electric shock, no modification of this equipment is allowed. Use only the included Ultradent VALO power supply and plug adapters. If these components are
damaged, do not use and call Ultradent Customer Service to order a replacement.
• Portable RF communications equipment may degrade performance if used closer than 30 cm (12 in").
• Use only authorized accessories, cables, and power supplies to prevent improper operation, increased electromagnetic emissions or decreased electromagnetic immunity (refer to
Electromagnetic Emissions section).
• To avoid the risk of electrical fire associated with handling of batteries:
o DO NOT mix rechargeable batteries with non-rechargeable batteries or other battery types.
o DO NOT attempt to charge non-rechargeable batteries.
o DO NOT autoclave or spray batteries, battery contacts, charger, or AC power adapter with liquid of any kind. If corrosion appears on the contacts of the battery charger,
call Ultradent Customer Service to order a replacement.
o DO NOT charge batteries around flammable materials.
o DO NOT keep charger in clinical operatory.
• To avoid the risk of injury, DO NOT use batteries that are corroded (rust), dented, emit an odor or fluids, have a torn or missing wrapping, or are otherwise damaged. Call Ultradent
Customer Service to order replacement batteries.
• To prevent the risk of thermal irritation or injury, avoid back-to-back curing cycles and do not expose oral soft tissues at close proximity for more than 10 seconds in any mode. If
longer curing times are required, use multiple shorter curing cycles or use a dual-cure product to avoid heating soft tissue.
• Use caution when treating patients who suffer from adverse photobiological reactions or sensitivities, patients who are undergoing chemotherapy treatment, or patients being
treated with photosensitizing medication.
• This unit may be susceptible to strong magnetic or static electric fields, which could disrupt the programming. If you suspect this has occurred, unplug the unit momentarily and then
re-plug it into the outlet.
VALO
Photo
Buttons and Lights
Power Button
Timing Lights
Low Battery - Indicator
Mode/Status Light
Timing/Mode
Change Button
Risk Group 2
Cordless
™
Curing Light
Legend
EN
3