Cleaning and Maintenance

WARNING
!
Beware of electrocution. Do not immerse the flow generator, humidifier or
power cord in water. Always unplug the device before cleaning and be sure
that it is dry before plugging it back in.
CAUTION
!
Do not use any of the following to clean any part of the humidifier. These
solutions may cause damage and reduce the life of the product:
- bleach, chlorine, alcohol- or ammonia-based solutions
- moisturising, antibacterial or glycerine-based soaps
- water softening and unapproved descaling agents.
Do not wash the water chamber in a dishwasher.
Daily
Wash the water chamber and connector seal in warm water, using a mild detergent
(see previous caution). Rinse them thoroughly with clean water and allow them to dry
away from direct sunlight.
Weekly
Inspect the water chamber for wear and deterioration. Replace the water chamber if
any component has cracked, become cloudy or pitted. Replace immediately if there is
any leakage from the water chamber.
If you notice white powder or deposits in the water chamber:
1. Fill the water chamber to the maximum water level mark with a solution of one part
vinegar to ten parts water.
2. Let the solution soak for ten minutes.
3. Discard the solution and rinse the water chamber thoroughly with clean water.
Periodically
Wipe the humidifier base with a clean cloth.