Low Level Disinfection (optional)
1. Disinfect the device in a disinfecting solution that has one of the following active Ingredients:
≥ 2.4% glutaraldehyde
or
>= 3.4% glutaraldehyde
or
≥ 0.55% ortho-phthalaldehyde
2. Prepare the disinfecting solution according to the manufacturer's instructions.
3. Per manufacturer's recommendations, immerse the device, filling all lumens, in the disinfecting
solution for the required time at the appropriate temperature.
4. Thoroughly rinse and flush all parts and lumens with running, demineralized water to remove
the disinfectant.
5. Dry all parts with a lint-free towel immediately after rinsing.
CIDEX Activated® is validated for disinfection efficacy.
4
CIDEX Plus® is validated for disinfection efficacy.
5
CIDEX® OPA is validated for disinfection efficacy.
6
Drying
For automated drying, use the drying cycle provided with the washer/disinfector.
For manual drying, use a lint-free cloth.
Dry any lumens with compressed air.
Maintenance, Inspection, and Testing
Inspect the device on a continual basis. If a problem is observed or suspected, the device
should be returned for repair.
Inspect all components for cleanliness. If fluid or tissue buildup is present, repeat the
above cleaning and disinfection procedures.
Inspect the camera cable for cuts and breaks. Return any damaged camera to Stryker for
service.
Packaging
N/A
Sterilization
After performing the cleaning instructions specified above, perform one of the following
sterilization cycles.
with a minimum soaking time of 45 minutes at 25 °C
4
with a minimum soaking time of 20 minutes at 25 °C
5
with a minimum soaking time of 12 minutes at 25 °C.
6
35