W A A R R N N I I N N G G / / C C A A U U T T I I O O N N S S
& & S S A A F F E E T T Y Y
Do not use Dopplers in the presence of
flammable gases such as anaesthetic
agents.
Do not use in the sterile field unless
additional barrier precautions are taken.
Do Not : Immerse in any liquid,
use solvent cleaner,
use high temperature sterilising
processes (such as autoclaving),
use E-beam or gamma radiation
sterilisation.
This product contains sensitive electronics,
therefore, strong radio frequency fields
could possibly interfere with it. This will be
indicated by unusual sounds from the
loudspeaker. We recommend that the
source of interference is identified and
eliminated.
Do not use in the eye.
If outputs 1 and 2 are used simultaneously,
the system should comply with EN60601-
1-1. Any equipment connected to outputs
1 or 2 should comply with EN60601-1,
EN60950, EN60065, EN60335 or
EN61010.
Do not dispose of batteries in fire as this
can cause them to explode.
Do not attempt to recharge normal dry-cell
batteries. They may leak, cause a fire or
even explode.
Dopplex Dopplers are screening tools to
aid the healthcare professional and should
not be used in place of normal vascular or
fetal monitoring. If there is doubt as to
vascularity or fetal well-being after using
the unit, further investigations should be
undertaken immediately using alternative
techniques.
We recommend that exposure to
ultrasound should be kept As Low As
Reasonably Achievable - (ALARA
guidelines). This is considered to be good
practice and should be observed at all
times.
Means Attention consult this
manual. Refer to safety section.
G G E E N N E E R R A A L L O O P P E E R R A A T T I I O O N N & &
M M A A I I N N T T E E N N A A N N C C E E
Refer to diagram on inside front cover for
Doppler Measuring sites and
Recommended Probes.
D D u u r r i i n n g g U U s s e e
An automatic noise reduction feature
operates on low level signals to improve
sound quality.
To replace battery see diagram.
A A f f t t e e r r U U s s e e
1.
Press and release the On/Off
button. If you forget to switch the
unit off, it will automatically shut-off
after 3 minutes.
2.
Refer to the cleaning section
before storing or using the unit on
another patient.
3.
Store unit together with probe and
accessories in the soft carry case
provided.
The control unit and the body of the probe
are robust and require no special handling.
However, the probe tip is delicate and must
be handled with care.
C C l l e e a a n n i i n n g g
Remove excess gel before parking the
probe. Clean with a damp cloth impregnated
with mild detergent. Do not allow fluid to
seep into either unit.
For disinfection use a soft cloth with Sodium
Hypochlorite 1000ppm or alcohol.
Alcohol saturated swabs can be used on the
probe tip.
Please be sure to check your local control of
infection policies, or any equipment cleaning
procedures.
Phenolic, detergent based disinfectants
containing cationic surfactants, ammonia
based compounds, or antiseptic solutions
such as Steriscol or Hibiscrub should never
be used on any part of the system.
C C o o u u p p l l i i n n g g G G e e l l
Use water based ultrasound gel ONLY.
3