17
Cleaning was
validated using
Enzol® at 1 oz/gal. at
35-40°C.
18
Prolystica 2X
Neutral Detergent at
1/8 oz/gal. at 35-40°C
is recommended for
cleaning.
Pre-Wash
Enzyme Wash
Wash 1
Rinse 1
Dry
19
Cleaning
was validated
using Neodisher
Mediclean Forte
TM
at 5 ml/L at 30°C.
Pre-Wash
Wash 1
Rinse 1
Dry
20
Drying was
validated with
pressurized air at
20 psi.
5. Automated Wash
•
Place the handpiece in an automated washer, on
an incline to aid in drainage.
•
Program the washer with the following
parameters, then activate the wash:
Recirculation
Time
2 Minutes
2 Minutes
2 Minutes
2 Minutes
7 Minutes
Recirculation
Time
2 Minutes
2 Minutes
2 Minutes
7 Minutes
6. Dry
Dry the handpiece using a clean cloth. Filtered
pressurized air can be used to assist drying.
7. Inspect
•
Visually inspect the handpiece, including all
internal surfaces, for remaining soil. Use an
endoscopic camera and endoscope if necessary
to see the inner surface of the lumen.
•
If soil remains, repeat the automated cleaning
procedure, focusing on those areas.
U.S. Parameters
Water
Temperature
Cold
Hot
60°C (set point)
Hot
115°C
EU Parameters
Water
Temperature
Cold
60°C (set point)
Hot
115°C
Detergent
Type
NA
Enzymatic
17
Non-Enzymatic
18
NA
NA
Detergent
Type
NA
Neodisher
Mediclean Forte
NA
NA
20
EN-25
19