Commissioning
3.5 Transducer Installation
10.Take either transducer and apply a continuous lengthwise 1/8-inch bead of coupling
11.Slide transducer into a Mounting Frames back end first aligning the angled edge of the
12.Tighten the transducer Clamping Screw to hold the transducer in place.
30
compound across the center of the transducer emitting surface.
①
F-Connector
②
Angled Edge
③
Transducer
④
Back End
⑤
Emitting Surface
⑥
Coupling Compound
Figure 3-6
Transducer Callouts
transducer with the angled edge of the Mounting Frame. Keep transducer from making
contact with the pipe until it butts up against the Mounting Frame stop.
procedure for the other transducer .
FUS1010 IP65 NEMA 4X Quick Start
Operating Instructions, 12/2008, CQO:QSG002 Revision B
Repeat installation