Design And Function - WIKA PG43SA-C Manual Del Usuario

Ocultar thumbs Ver también para PG43SA-C:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Supplementary documentation:
This additional information applies in conjunction with the
operating instructions "Pressure gauges for sanitary appli-
cations, models PG43SA-S, PG43SA-C" (article number
11532319).
This additional information supplements the respective
chapters of the above-mentioned document.

2. Design and function

2.3 Vent valve
Open
≤ 1 turn
By opening or closing the vent valve, the case will be vented to or
sealed from the atmosphere. Only ever operate the vent valve via the
spanner flats and to the prescribed torque using an appropriate tool.
By opening the vent valve, the pressure difference between the
atmosphere and the sealed case will be eliminated.
Maintaining the ingress protection is only ensured with a closed vent
valve.
WIKA additional information PG43SA-C, autoclavable version
Vent valve
Spanner width 7 mm
Close
Tightening torque:
≤ 0.35 Nm
EN
3
Tabla de contenido
loading

Tabla de contenido