Securing the transmitter:
in the connection head cover
①
Transmitter
③
Connection head
Note
Using the lock washers (TH200/TH300 only)
④
The lock washers
included in the delivery are only required for securely fastening the transmitter when the transmitter
is directly installed on the round plate
①
Fixing screw M4x35
③
Round plate
3.1.2
SITRANS TH installation on DIN rail and G rail
Note
Fixing rings
The fixing rings included in the scope of delivery for the transmitter are not required for the installation on DIN rails or G rails.
You can either install the transmitter on a 35 mm DIN rail or on a 32 mm G rail. DIN EN 60715 applies to DIN rails and G
rails in this context. The DIN rail adapter required for installation can be ordered as an accessory under the Order No.
7NG3092-8KA.
Adhere to the ambient conditions specified in the technical data.
8
③
for a temperature probe.
in the connection head base
②
Ceramic base of the measuring element
②
Transmitter
④
Lock washer DIN 6799 - 3.2 A2
②
SITRANS TH/TR/TF