Riester ri-pen Instrucciones Para El Uso página 26

Instrumentos diagnósticos
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 44
Emitted RF
IEC 61000-
4-3
Proximity
fields of
wire-
less RF
commu-
nications
equipment
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. The electroma-
gnetic propagation is affected by absorption and reflection by structures,
objects and people.
a: Field strengths of fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM broadcast and television transmission, cannot be theoretically
predicted with accuracy. To evaluate the electromagnetic environment
based on fixed RF transmitters, an electromagnetic assessment should be
considered. If the measured field strength at the location where the ri-pen
is used exceeds the above-mentioned RF degree of compliance, the ri-pen
should be observed to ensure normal operation.
If abnormal performance is observed, additional measures may be requi-
red, such as reorienting or moving the ri-pen.
b: With a frequency range of 150 kHz to 80 MHz, the field strengths should
be less than 3 V/m.
26
3 V/m
80 MHz to 2.7
GHz
380 – 390 MHz
27 V/m, PM 50%,
18 Hz
430 – 470 MHz
28 V/m; (FM ±5
kHz, 1 kHz sine)
PM, 18 Hz11
704 – 787 MHz
9 V/m, PM 50%,
217 Hz
800 – 960 MHz
28 V/m, PM 50%,
18 Hz
1700 – 1990 MHz
28 V/m, PM 50%,
217 Hz
2400 - 2570 MHz
28 V/m, PM 50%,
217 Hz
5100 - 5800 MHz
9 V/m, PM 50%,
217 Hz
10 V/m
27 V/m
28 V/m
9 V/m
28 V/m
28 V/m
9 V/m
Tabla de contenido
loading

Tabla de contenido