Guidelines and manufacturer's declaration – electromagnetic immunity
The ri-pen instrument is intended for use in the electromagnetic environ-
ment specified below. The customer or user of the ri-pen should ensure that
it is used in such an environment.
Immunity
IEC 60601
testing
test level
Conducted
3 Vrms
RF
0.5 MHz to 80
distur-
MHz
bances
6 V in ISM fre-
pursuant
quency bands
to
Between 0.15
IEC 61000-
MHz and 80 MHz
4-6
80% AM at 1 kHz
Compli-
Electromagnetic en-
ance
vironment - guidance
Not appli-
Portable and mobile RF
communications equip-
cable
ment should not be used
closer to any part of the
non-contact ri-pen, inclu-
ding the cables, than the
recommended distance,
which is calculated using
the equation applicable to
the transmitter frequency.
Recommended separation
distance:
d = 1.2 x P 80 MHz to
800 MHz
d = 2.3 x P 800 MHz to
2.7 GHz
Where P is the maximum
output power of the trans-
mitter in watts (W) accor-
ding to the transmitter
manufacturer and d is the
recommended distance in
metres (m).
Field strengths of fixed RF
transmitters, determined
by an electromagnetic
site survey, should be
lower than the compliance
standard in each frequency
range.
Interference may occur
in the vicinity of devices
marked with the following
symbol:
25