6556067_EN.book Seite 33 Montag, 18. April 2016 2:41 14
Operating Instructions BA ULTIMATE
7 Conditioning
3. Blow BA 695 / BA 688 out with max. 3 bar.
4. Maintain mechanical parts manually [ → 35].
5. Maintain the push button chuck manually [ → 38].
6. Pack BA 695 / BA 688 in packaging material suitable for sterilization
and storage. e.g. paper/plastic composite packaging.
7. Sterilize the adapter [ → 39].
7.3
Manual cleaning and disinfection
NOTICE! Condition immediately, or at the latest, one hour after
Immediately after treatment
treatment.
NOTICE! Never clean in an ultrasound bath!
NOTICE! Never immerse in disinfectant solution!
NOTICE! Using sprays from other manufacturers can reduce the
Using tbe BA Ultimate spray
product's service life. Only use BA Ultimate Spray.
IMPORTANT: Use a soft, clean, and disinfected brush for cleaning.
Use soft brush
33
65 56 067 D3560.201.09.02.02