6556067_EN.book Seite 37 Montag, 18. April 2016 2:41 14
Operating Instructions BA ULTIMATE
7 Conditioning
✔ The O-rings on the spray adapter are intact.
1. Fit the spray adapter onto the nozzle of the spray can.
2. Insert the turbine as far as it will go and hold it in place.
3. Spray into the turbine for 1-2 seconds.
IMPORTANT: Hold the spray can upright.
4. Wipe any spray that comes out with a disinfection cloth.
5. Repeat the process until the spray is clear.
37