Conditioning; Conduct Pre-Disinfection - B.A. International BA ULTIMATE Instrucciones De Uso

Ocultar thumbs Ver también para BA ULTIMATE:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 143
6556067_EN.book Seite 31 Montag, 18. April 2016 2:41 14
Operating Instructions BA ULTIMATE
65 56 067 D3560.201.09.02.02
7

Conditioning

7.1

Conduct pre-disinfection

✔ Wear appropriate protective clothing.
Predisinfection _ turbine
✔ All disinfectants must be approved in your country and have proven
bactericidal, fungicidal and virucidal properties. Only use
disinfectants with no protein-fixing properties.
1. Spray the surface with disinfectant.
2. Wipe the disinfectant off with a cloth.
 For further processing, the turbine should be dry and free of residue.
In the USA and Canada, for example, you can use:
Disinfectants in USA und Canada
®
● CAVICIDE
● CAVIWIPES
Please observe the manufacturer's instructions for using instrument
disinfectants.
7 Conditioning
31
Tabla de contenido
loading

Este manual también es adecuado para:

Ba 695Ba 688Ba695lw

Tabla de contenido