Closing the Polyclav
and pressurizing (Fig. 1+2)
®
1. Set the lid fully down on the bayonet catches on the vessel body. The marking line on the lid must be aligned
with the locking mechanism on the handle (Fig. 1).
2. Turn the locking lever to the "OPEN" ("AUF") position.
3. Rotate the lid until the two handles are in line (Fig. 2).
4. Turn the locking lever to the "CLOSED" ("ZU") position.
When the locking lever is in the "CLOSED" ("ZU") position:
a) the pressure release valve is closed, and
b) the lid is mechanically locked.
The lid cannot be removed when the Polyclav
Build up pressure in the vessel (Fig. 3)
1. Pressurize the closed vessel with the pressure hose to
2.0 – 2.2 bar (green marking line).
2. Should the pressure exceed the green marking line it can be reduced by briefly opening the locking lever.
3. Should the pressure exceed 2.5 bar (red marking line) the safety valve automatically opens and reduces the pressure to 2.0 bar.
Releasing the pressure and opening the Polyclav
1. Turn the locking lever to the "OPEN" ("AUF") position.
2. Wait until all the pressure has been released.
3. Turn the handle to the right until unlocked (the marking line on the lid should be aligned with the locking
mechanism on the handle) and then lift off.
When the locking lever is in the "OPEN" ("AUF") position
a) the pressure release valve is open and
b) the lid locking mechanism is released.
This ensures that the pressure is released before the lid can be opened.
 en
16
polyclav
®
is under pressure.
®
(Fig. 4)
®