Thermometer Display Icons and Alarms

Thermometer operation is communicated to the user via the LCD display and an audible device contained inside the thermometer
case. After the probe cover is installed or the batteries have been changed, the thermometer will perform a system reset. The
thermometer will perform an internal test to verify that the system components are functioning properly.
Alarm Condition
Patient temperature above specified range
Patient temperature below specified range
Ambient Temperature above specified range
Ambient Temperature below specified range
Low Battery
Dead Battery
If the display shows a system error 4 and the system error icon, the ambient conditions are changing too rapidly for the device to
be used. Allow the device to equilibrate for 20 minutes before using.
If the display shows any other system error, then reset the thermometer by installing a probe cover. If the system error does not
clear, then contact the factory for available service options. The applicable manufacturer addresses are located in Section XII,
Customer Service.
Tympanic Thermometer and Base
Display Mode
The LCD displays the low battery icon. The low battery icon will remain on until the
battery is replaced or until the dead battery icon is displayed. After the low battery
icon is displayed, the user may take approximately 100 temperatures until the dead
battery icon is displayed.
The LCD displays the dead battery condition. When any button is pressed, the
battery icon will flash several times and then the LCD will turn OFF. After the battery
icon is displayed, the batteries must be changed before the thermometer can
be used.
7
en