Covidien Genius 2 Manual De Instrucciones página 13

Ocultar thumbs Ver también para Genius 2:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 78
Displayed Temperature Measurement Range:
Temperature Range depends on Site Mode as follows:
Mode
Ear
Oral
Core
Rectal
Ambient Temperature Range:
16°C to 33°C (60.8°F to 91.4°F) where as CEN defines the standard as 16°C to 35°C (60.8°F to 95°F), 10 to 95 %RH, non-condensing.
Transport and Storage Temperature Range:
-25°C to 55°C (-13°F to 131°F), up to 95% RH non-condensing. If the unit is stored at extremes, it is recommended that the unit be
checked on the field calibration checker or at the factory before returning to service.
Clinical Repeatability:
Meets Section A.5 of EN 12470-5: 2003 (E) per Covidien llc technical report. Data is available from Covidien llc on request.
Response Time:
Less than 2 seconds
Pulse Timer:
60 seconds
Temperature Resolution:
0.1°C or 0.1°F
Power:
Internally Powered ME Equipment
3 AAA alkaline batteries
Battery Life:
Minimum of 15,000 temperature readings
Size:
Thermometer — 17.8 cm (7")
Base — 20.3 cm (8")
Weight:
Thermometer (with batteries) — 160 grams
Base — 100 grams
Degree of Protection Against Electrical Shock:
Type BF
Mode of Operation:
Continuous
Degree of Protection Against Ingress of Fluids:
Not Protected-IPX0
Tympanic Thermometer and Base
Range °C
33.0 to 42.0
33.6 to 42.0
34.0 to 42.0
34.2 to 42.0
Range °F
91.4 to 107.6
92.5 to 107.6
93.2 to 107.6
93.6 to 107.6
11