GAS AUTO-CHANGER (OPTIONAL)
5. Exchange the empty gas cylinder A for a new one immediately (Situation 5).
6. When cylinder B is empty, the supply reverts to cylinder A.
Table 4. CO
gas supply line automatic changeover
2
Situation
CO
gas is supplied
2
1
from valve A
CO
density in the
2
chamber is not
increased even
2
though the CO
gas
2
valve is open
CO
gas supply line
2
is automatically
3
changed to valve B
"Buzzer" is pressed
4
New cylinder A
5
fitted
Notes:
• When the Buzzer key is not pressed in Situation 4 and cylinder B becomes empty before cylinder A is
replaced, the supply will switch back to cylinder A. In this case, replace both cylinders and press "Buzzer"
immediately.
• The automatic changeover of CO
automatic changeover of cylinders may also occur in the situations where, e.g. the gas tube gets clogged,
gas is leaking, the gas pressure has dropped, the valve on the gas cylinder is insufficiently open etc, even
though the cylinder being used is not completely empty.
66
CO
gas
2
Supply line Cylinder A Cylinder B
A
Remaining Remaining
A
Empty
Remaining
B
Empty
Remaining
B
Empty
Remaining
B
Remaining Remaining
gas cylinders is linked to the CO
2
CO
gas supply line indicator
2
A
Light on
Light off
Light on
Light off
Reverse
video
Light on
and blink
Light off
Light on
Light off
Light on
density in the chamber. The
2
Message
display field
B
Err01:
CO
Gas Empty
2
(and buzzer)