1. Outer door: The outer door is held to the frame with the magnetic seal. The door heater is installed in
the door panel. The door opening is reversible. Contact our sales representative or agent to change the
door hinge from left to right or vice versa.
2. Gastight split door: Made of tempered glass, however avoid excessive impact on the glass.
3. Inner door: Open when take the trays out or clean inside.
4. Tray catches: Insert tray to fit the concave portion on the chamber.
5. Duct: The duct for the path for circulating air. It is removable.
6. Fan cover (above the duct): Serves as the inlet for circulating air. It is removable.
7. Fan (inside the duct): The fan is made from polypropylene resin. It can be disinfected in an autoclave.
1
8. UV lamp*
: This UV lamp does not generate ozone. For replacement, contact our sales representative
or agent.
9. Humidity control bar: Automatically reduces condensation caused by the external environment when
the door is frequently opened and closed. The humidity control bar has a bactericidal effect due to its
plated surface. However, it is recommended that it is replaced every 5 to 6 years to maintain this effect
(The duration of bactericidal effect differs depending on the environment).
10. Gas injection port: N
ON. For faster recovery of humidity in the chamber, connect the gas injection port and the gas injection
nozzle by the gas injection nozzle tube. Refer to page 19 for installing.
11. Humidifying pan cover: Prevents UV light entering the chamber.
12. Gas injection nozzle: Refer to page 19 for installing.
13. Humidifying pan: Fill with sterile distilled water.
14. Door switch: Detects the door opening/closing and stops the fan, electromagnetic valve for CO
electromagnetic valve for N
opening.
2
15. Key hole*
: This is unlocked with the unlock key when the outer door is locked by the electric lock.
16. Switch cover: Prevents shutting down of the unit in case of accidental pressing of the main power
switch.
17. Sample air outlet: This also functions as an internal gas outlet. In normal use, cover this outlet with
the sample air outlet cap.
18. Power switch: This is the main switch for the incubator (ON-"I", OFF-"O"). It also functions as an
overcurrent breaker.
3
19. Glow starter*
: Starts the glow for the UV lamp.
20. Connecting port A for CO
Note: When the optional MCO-21GC gas auto-changer is installed, both ports A and B are available. Refer
to page 65.
21. Connecting port A/B for N
22. Power supply cord cover plate: Prevents the power supply cord becoming detached.
23. USB port: Insert USB memory device to export operations and alarms log. Refer to pages 46–50.
gas (O
gas) is injected from here under the condition that O
2
2
/O
when the door is open. The UV lamp*
2
2
gas pipe: Refer to page 22 for gas cylinder connection.
2
/O
gas pipe: Refer to page 23 for gas cylinder connection.
2
2
density control is
2
1
is also deactivated by the door
, and