Software Configuration; Rotary Encoder Setting; Attaching The Trigger Electronics - Bosch CAM 847 Manual De Instrucciones

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

6.

Software configuration

i
If you are using the CRI 846/CRI 846H tester set, the
trigger electronics are already configured. In this case,
please skip this section.
Procedure:
1. Open the EPS945 system software.
 The software initializes the connected devices.
2. Select softkey <F7> in the "Main menu" dialog box.
1958
3. Enter the password:
4. Select the menu item "Components" submenu item
"Devices" in the "Configuration" dialog box and
open with <F12>.
5. In the "Devices" dialog box, select the "CRI800
injector control" field with u.
6. Open the list box with E .
7. In the list box, select the option "On" and accept
with E .
8. Press <F12> to confirm.
9. Exit the "Configuration" dialog box by pressing Esc.
 The system software initializes the connected de-
vices.
"
The status indicator for the CRI 800 trigger electronics
appears in the "main menu" dialog box.
Robert Bosch GmbH
and confirm with <F12>.
Software configuration | CAM 847 | 29
7.

Rotary encoder setting

7.1

Attaching the trigger electronics

Procedure:
1. Fasten the bracket to the trigger electronics.
Fig. 19:
Fastening the bracket
2. Hook the trigger electronics onto the oil pan of the
KMA.
Fig. 20:
Trigger electronics hooked in place
en
|
1 689 978 508
2011-12-03
Tabla de contenido
loading

Tabla de contenido