is set, the programmed levels are maintained over 24 hour. It is advisable to program at least two time periods to
differentiate day values and night values.
Attention: If two or more times for the start of temperature control are identical, the error E5 appears on the display.

5.3.3 Alarm setting temperature 2

When "00" flashes, use the ▲ and ▼ buttons to set the +/- alarm between 1 and 10° C. Press
.
"00" means that no alarm is set.

5.3.4 Deactivation/Activation of the audible temperature 2 alarm

When "OFF" is flashing, use the ▲ and ▼ buttons to activate the alarm with "ON" or to deactivate it by selecting
"OFF". Press
. When "TIME" is flashing.
If you want to finish programming, press and hold
for 3 seconds until the display returns to operating mode.
ATTENTION, pressing
only briefly will take you to the timer control setting (alternative option to temperature
control) for slot 2. The programmed values for temperature control are deleted the moment the program mode for
timer control is opened.