Operating Mode - Dentsply Sirona ProPex II Instrucciones De Uso

Ocultar thumbs Ver también para ProPex II:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

Procedure for lip clip, hook and fork
Operation
1.
Pre-Disinfection or
Decontamination
2.
Rinsing
3.
Manual Cleaning
4.
Rinsing
5.
Disinfection
6.
Final rinsing
7.
Inspection
8.
Packaging

Operating mode

- Soak immediately just after
usage all instruments in a
disinfectant solution
combined with proteolytic
enzyme if possible.
- Rinse manually and
abundantly the accessories
with current water
- clean manually the
accessories with an adequate
brush, preliminary soaked in
a clean pre-disinfectant
solution
See point # 2
- Immerse the accessories in a
disinfectant solution
(bactericidal, virucidal,
fungicidal, tuberculocidal and
aldehyde free) according to
the manufacturer
recommendations
- See point # 2
- After rinsing, the accessories
have to be dried.
- Inspect devices and sort out
those with defects.
- Pack the devices in
"Sterilization pouches".
Warning
- Follow instructions and observe
concentrations and immersion time
given by the manufacturer (an
excessive concentration may cause
corrosion or others defects on
instruments).
- The disinfectant solution should be
aldehyde free (to avoid blood
impurities fixation).
- Do not use disinfectant solution
containing Phenol or any products
which are not compatible with the
instruments (See general
recommendation).
- For visible impurities that are
observed on instruments, a pre-
cleaning is recommended by
brushing them manually with soft
material.
- the hook mechanism has to be
activated during the cleaning
process (press several times the
push button)
- no visible impurities should be
observed on the accessories
- Follow instructions and observe
concentrations and time given by
the manufacturer
- Dirty instruments must be cleaned
and disinfected again.
- Check the validity period of the
pouch given by the manufacturer to
X
X
X
X
X
X
X
X
14 / 71