6.6
Set the temperature to degrees Celsius (°C) or Fahrenheit (°F) in "temperature
setting".
6.7
The parameters for the measuring procedure are set in "measurement settings".
Fig. 6 - 1
For the fastest results (approx. 1 minute), set the parameter "meas. finished by:
predetermination". The DMA calculates the density before complete temperature
equilibrium has been reached.
For the highest accuracy results, set the parameter "meas. finished by:
equilibrium". The DMA determines the density and concentration after complete
temperature equilibrium has been reached.
In the DMA 5000 for higher accuracy 3 methods can be selected: Equilibrium fast,
Equilibrium medium, Equilibrium slow. The highest accuracy is achieved by using
Equilibrium slow.
XDLIB28
The factory setting for each method can be recalled by selecting "reset
method" in the menu "method settings".

Setting the Temperature

Settings for the Measuring Procedure

DMA 5000 screen: Measurement settings
"pump terminates measurement": If "yes" is selected, activating the air pump
will interrupt the measurement.
"pump switch off mode": If "time" is selected the air pump will be switched off
automatically after the specified "pump switch off time". Otherwise the air
pump has to be switched off manually.
Select the type of sample identification in "type of sample identif.".