Sony HVO-3300MT Manual Del Usario página 17

Ocultar thumbs Ver también para HVO-3300MT:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 56
Guidance and manufacturer's declaration – electromagnetic immunity
The HVO-3300MT is intended for use in the electromagnetic environment specified below. The customer or
the user of the HVO-3300MT should assure that it is used in such an environment.
IEC 60601
Immunity test
test level
Electrostatic
±6 kV contact
discharge (ESD)
IEC 61000-4-2
±8 kV air
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1 kV for input/
output lines
Surge
±1 kV line(s) to
line(s)
IEC 61000-4-5
±2 kV line(s) to
earth
Voltage dips,
< 5% U
short
(> 95% dip in
interruptions and
)
U
T
voltage
for 0.5 cycle
variations on
power supply
40% U
input lines
(60% dip in U
for 5 cycles
IEC 61000-4-11
70% U
(30% dip in U
for 25 cycles
< 5% U
(> 95% dip in
)
U
T
for 5 sec
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Compliance
level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV differential
mode
±2 kV common
mode
< 5% U
T
T
(> 95% dip in
)
U
T
for 0.5 cycle
40% U
T
T
)
(60% dip in U
T
for 5 cycles
70% U
T
T
)
(30% dip in U
T
for 25 cycles
< 5% U
T
T
(> 95% dip in
)
U
T
for 5 sec
3 A/m
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
HVO-3300MT requires continued operation during
power mains interruptions, it is recommended that
the HVO-3300MT be powered from an uninterruptible
power supply or a battery.
)
T
)
T
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
5
GB
Tabla de contenido
loading

Tabla de contenido