Inspection; Preventive Maintenance; Disposal Of The Medical Device; Disposing Of Non-Rechargeable Batteries - Dräger DigiFlow Instrucciones De Uso

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 50
Concept
Definition
Repair
Measures intended to restore the
functional condition of a medical de-
vice after a device malfunction

Inspection

Perform inspections at regular intervals and observe
the following specifications.
Checks
Interval
Personnel res-
ponsible
Inspection
Every
Service personnel
2 years
Inspection is no substitute for preventive
maintenance measures (including preventive
replacement of wear parts) as identified by the
manufacturer.
1
Check accompanying documents:
Instructions for use are available
2
Perform a functional test of the following features
according to the instructions for use:
Flow measurement
Internal battery
3
Check that the device combination is in good
condiction:
All labels are complete and legible
There is no visible damage

Preventive maintenance

Performing preventive maintenance
Replace battery.
Instructions for use DigiFlow

Disposal of the medical device

When disposing of the medical device:
Consult the relevant waste disposal company for
appropriate disposal.
Observe the applicable laws and regulations.
For countries subject to the EU
Directive 2002/96/EC
This device is subject to EU Directive 2002/96/EC
(WEEE). In order to comply with its registration
according to this directive, this device may not be
disposed of at municipal collection points for waste
electrical and electronic equipment. Dräger has
authorized a company to collect and dispose of this
device. To initiate collection or for further information,
visit Dräger on the Internet at www.draeger.com. Use
the Search function with the keyword "WEEE" to find
the relevant information. If access to Dräger's
website is not possible, contact the local Dräger
Organization.
Disposing of non-rechargeable
batteries
Do not recharge batteries.
Observe the applicable laws and regulations for
battery disposal.

Replacing the battery

low batt
The medical device does not contain any
repairable elements.
The medical device is calibrated at the factory.
Open the medical device only to the extent required
to replace the battery. Any attempt to further
dismantle the medical device will affect its accuracy.
If the display shows "low batt" or if the display does
no longer show any characters or symbols, the
battery must be replaced.
Use only the battery listed in the "Order list",
page 33. If other types of batteries are used, correct
functioning of the medical device cannot be
guaranteed.
Remove screws using a screwdriver (TORX, size
TX8).
Push the back cover backwards carefully.
English
27