6. To check flowmeter
the connecting tube
ing tube
off connection
17
eter and hold connection
closed. The flowmeter is free from leaks if
the flowmeter ball
the bottom.
7. Check mains failure alarm. To do this,
pull out the mains plug while the
OXYMAT 3 is switched on. An acoustic
signal is heard, and the red fault
indicator
lights up.
18
8. Insert the mains plug in the power socket
again. The alarm signal should stop, and
the device starts again.
9. Switch off your OXYMAT 3 when you no
longer need it.