7. Functional Check
1. Switch on the OXYMAT 3 using On/Off
knob
. A self-test will then run.
2
– Red fault indicator
dicator
and green power indicator
19
immediately come on briefly.
20
– Green power indicator
es for the duration of the self-test (ap-
prox. 1 minute).
– If the device is functioning perfectly,
only green power indicator
be on after the self-test.
– Yellow status indicator
ing of a fault. You should then have
the device checked by your author-
ized dealer.
– Red fault indicator
vere fault. The OXYMAT 3 should be
taken out of service.
2. Check that the red minutes counter on the
operating time meter
6 minutes.
3. Check that the screw connections be-
tween the cover and glass of the bubble
humidifier and between the humidifier
and dosage monitor are firm. If neces-
sary, tighten these connections by hand.
4. Set the flow rate to 2 liters per minute by
means of the oxygen dosage control
knob
.
24
5. To check the dosage monitor
bend 3-meter tube
3-meter hose
11
bubble humidifier
nection
closed. The dosage monitor
26
system is free from leaks if the flowmeter
ball
sinks all the way to the bottom. In
22
the event of leaks you should replace the
sealing rings on the bubble humidifier.