General Technical Specifications - ResMed S9 AutoSet Manual De Instrucciones

Ocultar thumbs Ver también para S9 AutoSet:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Problem/Possible Cause
Displays error message: Tube blocked, please check your tube
Air tubing is blocked.
The following message is displayed on the LCD after you try to update settings or copy data to the SD card: Card
error, please remove SD card and contact service provider
SD card is not inserted correctly.
You may have removed the SD card before
settings were copied to the CPAP device.
The following message is NOT displayed on the LCD after you try to update the settings using the SD card: Settings
updated successfully, press any key
The settings were not updated.
Water splashing on your face from the H5i
The water tub is overfilled.
Condensation is forming in the air tube and mask.
Leaking water tub
The water tub may be damaged or cracked.
The cleanable water tub is not assembled
correctly.
Air feels too warm/cold in the mask
The temperature of the ClimateLine air tubing is
set too high/low.

General technical specifications

90W power supply unit
30W power supply unit
Operating temperature
Operating humidity
Operating altitude
Storage and transport temperature
Storage and transport humidity
4
Solution
Check your air tubing and remove any blockages.
Ensure that the SD card is inserted correctly.
Reinsert the SD card and wait for the home screen or the Settings
updated successfully, press any key message to appear on
the LCD.
Note: This message only appears once. If you re-insert the SD card after
you have updated your settings, the message will not be re-displayed.
Contact your clinician/service provider immediately.
Check that the water level is below the maximum water level mark.
Turn the H5i or relative humidity setting down via the menu.
Contact your service provider for a replacement.
Check for damage and reassemble the cleanable water tub correctly.
Turn up/down the heated tubing temperature via the menu.
Input range 100–240V, 50–60Hz,
Nominal for aircraft use 110V, 400Hz
Typical power consumption 70W (80VA)
Maximum power consumption 110W (120VA)
Input range 100–240V, 50–60Hz,
Nominal for aircraft use 110V, 400Hz
Typical power consumption 20W (30VA)
Maximum power consumption 36W (75VA)
º
º
º
41
F to 95
F (+5
C to +35
10–95% non-condensing
Sea level to 8,500' (2,591 m)
º
º
º
-4
F to 140
F (-20
C to +60
10–95% non-condensing
º
C)
º
C)