our Technical Service Department at 1-800-344-3100 for a return authorization
number or visit www.lamotte.com for troubleshooting help. The sender is
responsible for shipping charges, freight, insurance and proper packaging to
prevent damage in transit. This warranty does not apply to defects resulting
from action of the user such as misuse, improper wiring, operation outside of
specifi cation, improper maintenance or repair, or unauthorized modifi cation.
LaMotte Company specifi cally disclaims any implied warranties or merchantability
or fi tness for a specifi c purpose and will not be liable for any direct, indirect,
incidental or consequential damages. LaMotte Company's total liability is limited to
repair or replacement of the product. The warranty set forth above is inclusive and
no other warranty, whether written or oral, is expressed or implied.
REGISTER YOUR METER
To register your meter with the LaMotte Service Department, go to
www.lamotte.com and choose SUPPORT on the top navigation bar.
TROUBLESHOOTING
ERROR MESSAGES
Over Range
If the message OVERRANGE is displayed when scanning a sample, the sample
may be over range or under range. If the sample is over range the sample should
be diluted and tested again (see Sample Dilution Techniques and Volumetric
Measurements, page 21).
If overrange is displayed, press
to continue testing on diluted samples.
Note: After pressing
overrange cncentration will be
displayed. This concentration is an
approximation only.
CALIBRATION
As with all pre-calibrated meters, it is highly recommended, even if not required
by regulations, that the user periodically verify the performance of the meter
by running standards with a predetermined concentration. Results outside of
specifi cation are an indication that the meter needs to be adjusted. This can be
done following the user calibration described on page 9. If the user calibration fails
to properly adjust the meter then the meter should be returned to LaMotte Company
for recalibration. (See page 23).
STRAY LIGHT
The colorimeter should have no problems with stray light. Make sure that the
sample compartment lid is always fully closed.