General Information; Packaging And Delivery; General Precautions; Safety Precautions - LaMotte DC1500 Manual De Instrucciones

Ocultar thumbs Ver también para DC1500:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Amount of Sample
10 mL
5 mL
2.5 mL
1 mL
0.5 mL
All dilutions are based on a fi nal volume of 10 mL, so several dilutions will require
small volumes of the water sample. Graduated pipets should be used for all
dilutions. If volumetric glassware is not available, dilutions can be made with the
colorimeter tube. Fill the tube to the 10 mL line with the sample and then transfer
it to another container. Add 10 mL volumes of deionized water to the container
and mix. Transfer 10 mL of the diluted sample to the colorimeter tube and follow
the test procedure. Repeat the dilution and testing procedures until the result falls
within the range of the calibration. Multiply the test result by the dilution factor.
For example, if 10 mL of the sample water is diluted with three 10 mL volumes of
deionized water, the dilution factor is four. The test result of the diluted sample
should be multiplied by four.

GENERAL INFORMATION

 PACKAGING AND DELIVERY
Experienced packaging personnel at LaMotte Company assure adequate protection
against normal hazards encountered in transportation of shipments.
After the product leaves LaMotte Company, all responsibility for safe delivery is
assured by the transportation company. Damage claims must be fi led immediately
with the transportation company to receive compensation for damaged goods.
 GENERAL PRECAUTIONS
READ THE INSTRUCTION MANUAL BEFORE ATTEMPTING TO SET UP OR OPERATE THE
METER. Failure to do so could result in personal injury or damage to the meter. The
meter should not be used or stored in a wet or corrosive environment. Care should
be taken to prevent water from wet tubes from entering the meter chamber.
NEVER PUT WET TUBES IN THE METER.
 SAFETY PRECAUTIONS
Read the labels on all LaMotte reagent containers prior to use. Some containers
include precautionary notices and fi rst aid information. Certain reagents are
considered hazardous substances and are designated with a * in the instruction
manual. Safety Data Sheets (SDS) can be found at www.lamotte.com. Read the
SDS before using these reagents. Additional emergency information for all LaMotte
reagents is available 24 hours a day from the National Poison Control Center 1-800-
222-1222 or by contacting the 24 hour emergency line for ChemTel 1-800-255-
3924 (USA, Canada, Puerto Rico); locations outside the North American Continent
813-248-0585 (call collect). Be prepared to supply the name and four-digit
LaMotte code number found on the container label or at the top of the SDS or in the
contents list of the procedure. LaMotte reagents are registered with a computerized
poison control information system available to all local poison control centers.
Keep equipment and reagent chemicals out of the reach of young children.
Ensure that the protection provided by this equipment is not impaired. Do not intall
or use this equipment in a manner that is not indicated in this manual.
Deionized Water to Bring
Final Volume to 10 mL
0 mL
5 mL
7.5 mL
9 mL
9.5 mL
21
Multiplication Factor
1
2
4
10
20
Tabla de contenido
loading

Este manual también es adecuado para:

3240