High-Level Disinfection Of The Prismatic Tube Without Sapphire Glass - Dentsply Sirona CEREC Bluecam Manual De Instrucciones

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

3 CEREC Bluecam
3.1 CEREC Bluecam without sapphire glass
18
3.1.7
High-level disinfection of the prismatic tube without
sapphire glass
NOTE
Hot air sterilization and high-level disinfection must not be combined.
The complete process for high-level disinfection (HLD) is as follows –
assuming the CIDEX
your dealer:
A HLD set to support the HLD processes can be ordered from Dentsply
Sirona with REF 63 46 907.
1. Remove the camera support (if present).
2. Carry out a preliminary cleaning process (see "Pre-cleaning of the
prismatic tube  [→ 17]").
3. Always use personal protective equipment when implementing the
high-level disinfection if not before.
4. Remove the prismatic tube from the CEREC Bluecam (see
"Removing the prismatic tube  [→ 16]").
5. Place the protective cap on the camera head and place the camera
in the camera storage unit.
6. Use the following disinfectants for the high-level disinfection:
®
CIDEX
OPA.
CAUTION
Observe the safety instructions from the disinfectant manufacturer.
Instructions for CEREC Bluecam and CEREC Omnicam
®
OPA disinfectant is available in your country via
Dentsply Sirona