2 CEREC Omnicam
2.8 High-level disinfection of the mirror sleeve with the HLD set 63 46 907
12
Instructions for CEREC Bluecam and CEREC Omnicam
6. Press the green plug into the interior of the mirror sleeve in order to
ensure that no particle contamination or fluids penetrate the inside
of the mirror sleeve.
NOTE
Nature and source of the danger
The plug does not seal the mirror sleeve watertight.
7. Carefully fill the HLD container up to the 50 ml mark ("Omnicam"
mark). A funnel can be used to help pour in the liquid. Do not spill
the disinfectant. Should you spill any disinfectant, follow the safety
instructions from the disinfectant manufacturer.
NOTE
Nature and source of the danger
Only place the mirror sleeve into the container and not the entire
camera.
Ensure that the mirror sleeve is placed upright in the disinfectant (with
the coated sapphire glass down) so that the solution cannot penetrate
the inside of the mirror sleeve.
8. Insert the mirror sleeve in the correct position (coated sapphire
glass down).
9. Place the cover on the container and leave the mirror sleeve in the
disinfectant for at least 12 minutes (CIDEX
will not be damaged if it remains in the disinfectant for longer than
the required time, but removing it shortly after the required time is
recommended.
10. Remove the mirror sleeve from the container.
11. Carefully rinse the mirror sleeve with tap water for at least 30
seconds. Ensure that no water is able to penetrate into the interior
of the mirror sleeve.
12. Wipe down the mirror sleeve. Use cotton wool gauze to do so,
moistened with tap water.
13. Dry the mirror sleeve using a soft, lint-free cloth.
14. Remove the green plug from the mirror sleeve.
Dentsply Sirona
®
OPA). The mirror sleeve