ResMed VPAP Malibu Manual De Usuario página 37

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

268106r1.book Page 32 Thursday, November 30, 2006 2:35 PM
Problem
VPAP Malibu unit does
not stop when you
remove your mask.
SmartStop is enabled
but the flow generator
does not stop
automatically when
you remove your mask.
Display error message:
Check tube!!
Key if done
32
Possible Cause
There is excessive leak.
Mask Alarm has been
enabled; SmartStop has
automatically been
disabled.
Use of a full face mask.
Incompatible humidifier or
mask system being used.
The air tubing is loose.
Air tubing is blocked.
The device motor may not
be running at the correct
speed..
Solution
Adjust position of mask and
headgear.
Plugs may be missing from
ports on mask. Replace
them.
Air tubing not connected
properly. Connect firmly at
both ends.
Air tubing kinked or
punctured. Straighten or
replace.
Disable Mask Alarm to
enable SmartStop.
SmartStop does not work
with a full face mask.
Use only equipment as
recommended and supplied
by ResMed.
Check that the air tubing is
connected securely to your
mask and the air outlet on
the front of the
VPAP Malibu.
Check air tubing for
blockage.
Return your unit for
servicing.
To clear the error message,
press any key on the
VPAP Malibu keypad.
Tabla de contenido

Solución de problemas

loading

Tabla de contenido