Zero Calibration - MSA ALTAIR 5 Manual De Instrucciones

Ocultar thumbs Ver también para ALTAIR 5:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

OPERATION

3.9.1 Zero calibration

(1) Press the
Measurement page.
ZERO screen displays.
To skip the ZERO procedure and move
directly to the calibration span
procedure, push the
button is pushed for 30 seconds, the
instrument prompts user to perform a
SPAN calibration before instrument
returns to the Normal Measurement
page.
(2) Press the
ZERO screen, i. e. to execute zero
calibration.
LEDs flash
"
ZERO CALIBRATION
blinks
ZERO calibration starts.
A progress bar shows the
user how much of the
calibration has been
completed.
After the ZERO calibration is
completed the instrument
displays either
"
ZERO CALIBRATION
PASS
or
"
ZERO CALIBRATION
FAIL
Only if the instrument passes
the zero calibration the SPAN
screen displays.
40
button in Normal
button. If no
button to confirm the
"
".
ALTAIR 5 - Operating Manual
"
MSA
1:30 PM
SETUP
%
%
0
20.8
O
COMB
2
PPM
PPM
CO
H S
0
0
2
SO
2
PPM
0
ZERO CALIBRATION ?
YES
NO
1:30 PM
SETUP
%
%
0
20.8
O
COMB
2
PPM
PPM
0
CO
H S
0
2
Cl
2
PPM
0.0
ZERO CALIBRATION
65%
1:30 PM
SETUP
%
%
O
COMB
2
PPM
PPM
H S
CO
2
Cl
PPM
2
ZERO CALIBRATION PASS
1:30 PM
SETUP
%
O
2
PPM
Cl
2
ZERO CALIBRATION FAIL
1:30 PM
SETUP
%
%
58
15.0
O
COMB
2
PPM
PPM
CO
H S
20
60
2
SPAN CALIBRATION ?
YES
NO
US