OPERATION
3.9.3 Finishing Calibration
(1) Close the valve on the regulator.
(2) Remove the calibration cap or tube from pump inlet.
The calibration procedure adjusts the span value for any sensor that passes
the calibration test. Sensors that fail calibration are left unchanged.
Since residual gas may be present, the instrument may briefly go into an
exposure alarm after the calibration sequence is completed.
3.9.4 Autocalibration Failure
If the instrument cannot calibrate one or more sensor(s), it goes into the
Calibration Failure page and remains in alarm for 10 seconds.
Sensors that could not be calibrated are indicated by dashed lines on the
concentration display.
If a sensor repeatedly fails calibration after the full calibration procedure in
this manual has been performed, replace the combustible sensor.

3.10. Instrument Shutdown

For instrument shutdown press and hold the
Figure 3-10 Instrument Shutdown
The instrument displays a blinking "
SHUTDOWN
hold the button to complete the shutdown.
42
" and a progress bar shows the user how much longer to
ALTAIR 5 - Operating Manual
button.