Error Message
T r a n s
There is a fault with the conductivity sensor.
T r a n s
There is a fault with the pressure transducer
or wiring.
There is a fault with the water pump.
The calibration file is out of date.
g
Internal software time out.
There is a fault with the float chamber water
level transducer.
B
E R
There is a fault in the float chamber; the
water level is too high.
R e q
d
A service is required.
The oxygen vent is blocked.
S t o p
The terminals on JP3 of the options board
are short circuit.
Pressure error, O
consecutive errors of the same type have
occurred.
H2 Production
(System locked)
vent blocked or three
2
(System locked)
Action
No
Contact Parker Hannifin for advice.
Yes
Perform the required service.
Remove the blockage and perform a
hard reset.
No
If the problem persists contact Parker
Hannifin for advice.
Press
when required to reset
No
the generator.
No
Resolve the initial fault and reset the
generator.