Battery Replacement; Service And Replacement Parts - Ideal 61-796 Manual De Instrucciones

Ocultar thumbs Ver también para 61-796:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

*
Follow the proper connection as shown in Fig. 1. The LED (red) indicator
will light. This indicates that a correct current flow is taking place.

Battery replacement:

When the symbol "
Disconnect the test leads from the instrument and turn off the power.
Unscrew the cover, then slide the cover, take out the batteries and replace with new
batteries, type AA.
Replace back cover and secure with 2 screws.
If the meter is not to be used for periods of longer than 60 days, remove the batteries
and store them separately.
Maintenance:
Clean the case with a damp cloth and mild detergent : do not use abrasives or solvents.

Service and Replacement Parts:

This unit has no user-serviceable parts.
For replacement parts or to inquire about service information, contact IDEAL INDUSTRIES, INC.
at 1-877-201-9005 or visit our website www.testersandmeters.com.
SPECIFICATIONS
Measurement System:
Earth resistance:
Resolution:
Earth voltage:
Accuracy:
Earth voltage:
Display:
Low battery indication:
Data hold indication:
Over range indication:
Safety standard:
Power source:
Dimensions:
Weight:
Accessories included:
" appears on the display, replace with new batteries as follows:
Operating Frequency - 820Hz, 2mA approx.
Range: 20/200/2000Ω
0~19.99Ω (0.01Ω)
0~199.9Ω (0.1Ω)
0~1999Ω (1Ω)
0~200V AC, 40-500Hz
Earth resistance: ±(2%reading± 2digits) or ±0.1Ω,
whichever is greater.
±(1% reading±2digits)
LCD 3½ digit (2000 counts)
"
"Symbol appears on the display.
"
HOLD
" Symbol appears on the display.
" 1 " (MSD).
IEC/EN 61010-1, Installation CATIII - 200V.
(8) 1.5V (AA) batteries required.
8.3"Hx8.3"Wx4.0"D (210mmLx210mmWx100mmD)
Approx. 3 Lbs(1361g) (batteries included).
Hard Case, Test leads (red-33m, yellow-33m, green-5m), 2 Auxiliary
earth spikes, simplified measurement probe, Instruction manual.
3
Tabla de contenido
loading

Tabla de contenido